Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8000A: (1 year) CAL VER /5100,8842 DATE: 08-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 34 NUMBER OF LINES: 162 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 8842A ============================================================================= # # Source: # Fluke 8000A instruction manual (PN 347906). # # Compatibility: # MET/CAL 3.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Switch the Unit Under Test (UUT) On 1.005 DISP and ensure that it has warmed up. 1.006 ASK- N U 1.007 HEAD RESISTANCE TESTS (Test Lead Compensation) 1.008 DISP Configure the 8842A as follows: 1.008 DISP [32] 8842A INPUT HI to 8842A SENSE HI 1.008 DISP [32] 8842A INPUT LO to 8842A SENSE LO 1.009 IEEE [@8842]G5[I] 1.010 MATH MEM1 = MEM - 1100 1.011 JMPF 1.014 1.012 DISP Select FRONT INPUTS on the 8842A. 1.013 JMP 1.009 1.014 DISP Connect the 5100B OUTPUT to the 8842A INPUT. 1.015 RESF 100Z S 2W 1.016 8842 100Z N 4W 1.017 MATH M[1] = MEM 1.018 RESF 1kZ S 2W 1.019 8842 1kZ N 4W 1.020 MATH M[2] = MEM 1.021 RESF 10kZ S 2W 1.022 8842 10kZ N 4W 1.023 MATH M[3] = MEM 1.024 RESF * S 1.025 HEAD {RESISTANCE TESTS} 1.026 DISP Disconnect the leads from the 8842A and connect them to 1.026 DISP the UUT V-Ohm and COMMON terminals. 1.027 HEAD RESISTANCE TESTS: {200 Ohm Range} 1.028 RNG 200Z 1.029 RESF 100.0Z S 2W 1.030 ASK+ N U 1.031 MATH MEM = M[1] # 8842A 90 day accuracy, 200 Ohm Range 1.032 ACC 100 Z 0.007% 0.004U 1.033 MEMI Enter UUT reading in Ohms: 1.034 MEME 1.035 MEMC 200 Z 0.05/ 0.2% 2.001 ASK- U 2.002 HEAD RESISTANCE TESTS: {2 kOhm Range} 2.003 RNG 2kZ 2.004 RESF 1.000kZ S 2W 2.005 ASK+ U 2.006 MATH MEM = M[2] # 8842A 90 day accuracy, 2 kOhm Range 2.007 ACC 1 kZ .005% .00003U 2.008 MEMI Enter UUT reading in kilohms: 2.009 MEME 2.010 MEMC 2 kZ 0.05/ 0.2% 3.001 ASK- U 3.002 HEAD RESISTANCE TESTS: {20 kOhm Range} 3.003 RNG 20kZ 3.004 RESF 10.00kZ S 2W 3.005 ASK+ U 3.006 MATH MEM = M[3] # 8842A 90 day accuracy, 20 kOhm Range 3.007 ACC 10 kZ .005% .0003U 3.008 MEMI Enter UUT reading in kilohms: 3.009 MEME 3.010 MEMC 20 kZ 0.05/ 0.2% 4.001 HEAD RESISTANCE TESTS: {200 kOhm Range} 4.002 RESF 200 100.0kZ 0.05/ 0.2% 2W 5.001 HEAD RESISTANCE TESTS: {2000 kOhm Range} 5.002 RESF 2000 1000kZ 0.05/ 0.2% 2W 6.001 HEAD RESISTANCE TESTS: {20 MOhm Range} 6.002 RESF 20 10.00MZ 0.05/ 0.5% 2W 7.001 HEAD {DC VOLTS AND LINEARITY TESTS} 7.002 HEAD DC VOLTS AND LINEARITY TESTS: {200mV Range} 7.003 5100 200 190.0mV 0.05/ 0.1% 2W 8.001 5100 200 -190.0mV 0.05/ 0.1% 2W 9.001 HEAD DC VOLTS AND LINEARITY TESTS: {2V Range} 9.002 5100 2 1.900V 0.05/ 0.1% 2W 10.001 HEAD DC VOLTS AND LINEARITY TESTS: {20V Range} 10.002 5100 20 1.000V 0.05/ 0.1% 2W 11.001 5100 20 10.00V 0.05/ 0.1% 2W 12.001 5100 20 19.00V 0.05/ 0.1% 2W 13.001 HEAD DC VOLTS AND LINEARITY TESTS: {200V Range} 13.002 5100 200 190.0V 0.05/ 0.1% 2W 14.001 HEAD DC VOLTS AND LINEARITY TESTS: {1200V Range} 14.002 5100 1200 1000V 0.08/ 0.1% 2W 15.001 HEAD {AC VOLTS TESTS} 15.002 HEAD AC VOLTS TESTS: {1200V Range} 15.003 5100 1200 1000V 0.17/ 0.5% 400H 2W 16.001 HEAD AC VOLTS TESTS: {200V Range} 16.002 5100 200 190.0V 0.1/ 0.5% 400H 2W 17.001 HEAD AC VOLTS TESTS: {20V Range} 17.002 5100 20 19.00V 0.1/ 0.5% 400H 2W 18.001 HEAD AC VOLTS TESTS: {2V Range} 18.002 5100 2 1.900V 0.1/ 0.5% 400H 2W 19.001 HEAD AC VOLTS TESTS: {200mV Range} 19.002 5100 200 190.0mV 0.1/ 0.5% 400H 2W 20.001 HEAD {FREQUENCY RESPONSE TESTS} 20.002 5100 200 190.0mV 0.1/ 0.5% 50H 2W 21.001 5100 200 190.0mV 0.1/ 0.5% 10kH 2W 22.001 5100 200 190.0mV 0.1/ 1% 20kH 2W 23.001 HEAD {DC CURRENT TESTS} 23.002 DISP Connect the 5100B and UUT as follows: 23.002 DISP [32] 5100B OUTPUT HI to UUT mA 23.002 DISP [32] 5100B OUTPUT LO to UUT COMMON 23.003 HEAD DC CURRENT TESTS: {200uA Range} 23.004 5100 200 190.0uA 0.05/ 0.3% 2W 24.001 5100 200 -190.0uA 0.05/ 0.3% 2W 25.001 HEAD DC CURRENT TESTS: {2mA Range} 25.002 5100 2 1.900mA 0.05/ 0.3% 2W 26.001 HEAD DC CURRENT TESTS: {20mA Range} 26.002 5100 20 19.00mA 0.05/ 0.3% 2W 27.001 HEAD DC CURRENT TESTS: {200mA Range} 27.002 5100 200 190.0mA 0.05/ 0.3% 2W 28.001 HEAD DC CURRENT TESTS: {2000mA Range} 28.002 5100 2000 1900mA 0.05/ 0.3% 2W 29.001 HEAD {AC CURRENT TESTS} 29.002 HEAD AC CURRENT TESTS: {2000mA Range} 29.003 5100 2000 1900mA 0.1/ 1% 400H 2W 30.001 HEAD AC CURRENT TESTS: {200mA Range} 30.002 5100 200 190.0mA 0.1/ 1% 400H 2W 31.001 HEAD AC CURRENT TESTS: {20mA Range} 31.002 5100 20 19.00mA 0.1/ 1% 400H 2W 32.001 HEAD AC CURRENT TESTS: {2mA Range} 32.002 5100 2 1.900mA 0.1/ 1% 400H 2W 33.001 HEAD AC CURRENT TESTS: {200uA Range} 33.002 5100 200 190.0uA 0.1/ 1% 400H 2W 34.001 END